Use of a Pressure Cooker to Achieve Sterilization for an Expeditionary Environment

Top

Cook RK, McDaniel J, Pelaez M, Beltran T, Webb O. 21(1). 37 - 39. (Journal Article)

Abstract

Background: Sterilization of healthcare instruments in an expeditionary environment presents a myriad of challenges including portability, cost, and sufficient electrical power. Using pressure cookers to sterilize instruments presents a low-cost option for sterilization in prehospital settings. This project's objective was to determine if sterility can be achieved using a commercially available pressure cooker. Methods: Presto® 4-quart stainless steel pressure cookers were heated using Cuisinart® CB-30 cast-iron single burners. One 3M™ Attest™ 1292 Rapid Readout Biological Indicator and one 3M™ Comply™ SteriGage™ integrator strip were sealed in a Henry Schein® Sterilization Pouch and placed in a pressure cooker and brought to a pressure of 103.4kPa. Sterility was verified after 20 minutes at pressure. The Attest vials were incubated in a 3M Attest 290 Auto-Reader for 3 hours with a control vial. Results: Sterility using the pressure cooker was achieved in all tested bags, integrator strips, and Attest vials (n = 128). The mean time to achieve the necessary 103.4kPa was 379 seconds (standard deviation (SD) = 77). Neither the ambient temperature nor humidity were found to affect the pressure cooker's time to achieve adequate pressure, nor the achieved depth on the integrator strip (all p > .05). Conclusion: This study provides evidence that sterilization is possible with offthe- shelf pressure cookers. Though lacking US Food and Drug Administration (FDA) approval, the use of this commercially available pressure cooker may provide a method of sterilization requiring minimal resources from providers working in expeditionary environments.

Keywords: disinfection; sterilization; dental instruments; health professional

PMID: 33721304

DOI: WPGC-A599

Buy Now

Watch Where You Point That: Pneumomediastinum From Pneumatic Nail Gun Injury to the Hand

Top

Nam JJ, Kelly WF. 21(1). 106 - 108. (Journal Article)

Abstract

Pneumatic nail guns are hand-held tools used in industrial and construction settings. Nail guns cause the most trauma with hospitalization among construction workers. To our knowledge, we report for the first time a case of pneumomediastinum from a nail gun injury to the hand. Our patient was a 40-year-old male construction worker who shot a nail gun into his hand. He became acutely dyspneic and was found to have a pneumomediastinum due to air insufflation. He later underwent tube thoracostomy and intubation. To our knowledge, this is the first report of pneumomediastinum from a nail gun injury to the hand.

Keywords: pneumatic nail guns; pneumomediastinum; nail gun injury

PMID: 33721316

DOI: 7R9Z-U2IE

Buy Now